MAXIMO II CRT-D
Report
- Report Number
- 3004209178-2013-01762
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 16, 2012
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 6947, IMPLANTABLE TACHY LEAD, (B)(6) 2009; 4076, IMPLANTABLE PACING LEAD, (B)(6) 2009. (B)(4).
IT WAS REPORTED THAT THE DEVICE WAS AT ELECTIVE REPLACEMENT INDICATOR (ERI) AND HAD EARLY BATTERY DEPLETION. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD MOVED FROM ITS LOCATION TO AN AREA WHERE THE LEAD WOULD NOT WORK. THE DEVICE WAS EXPLANTED AND REPLACED AND THE LEAD WAS CAPPED AND NOT REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55882 | MAXIMO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D284TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00051 YR | Hospitalization| R | 4196 IMPLANTABLE PACING LEAD |