DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE
Report
- Report Number
- 2649622-2013-00662
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 16, 2012
- Report Date
- February 21, 2013
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS OVER-ROTATED, THERE WAS BLOOD ON THE DISTAL ELECTRODE, AND THE LEAD WAS DAMAGED AT IMPLANT. THE ANALYST NOTED THAT THE LEAD WAS RETURNED WITH THE DISTAL CONDUCTOR DISTORTED WITHIN THE CONNECTOR. THE DISTAL CONDUCTOR HAS BEEN OVER-ROTATED. WHEN THE HELIX WAS RETRACTED DURING THE PROCEDURE, THE IS-1 CONNECTOR PIN WAS TURNED AN EXCESSIVE NUMBER OF TIMES. THIS PREVENTS PROPER TORQUE TRANSFER WHEN TURNING THE IS-1 PIN.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT WHEN THE LEAD WAS PLACED DURING THE IMPLANT PROCEDURE THE HELIX OF THE LEAD WOULD NOT MOVE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57655 | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00085 YR |