CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2013-00642
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 19, 2012
- Report Date
- January 30, 2013
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND NO ANOMALIES. HOWEVER, IT WAS NOTED THAT THE DISTAL ELECTRODE WAS COVERED IN BLOOD. VISUAL ANALYSIS NOTED THAT THE LEAD WAS RECEIVED WITH STYLET IN THE LEAD AND THE HELIX BENT. THE BENT HELIX IS LIKELY DUE TO ATTEMPTED IMPLANT DAMAGE. DUE TO THE HELIX BEING BENT, THE EXTENSION/RETRACTION AND LENGTH TESTING COULD NOT BE PERFORMED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT AT THE IMPLANT PROCEDURE, THE RIGHT VENTRICULAR (RV) LEAD WAS POSITIONED AND WHEN THE ATRIAL LEAD WAS ATTEMPTED TO BE PLACED IT WAS VERY DIFFICULT TO MOVE. THERE WAS A REMARKABLE COMPRESSION OF ONE LEAD AGAINST THE OTHER WHICH MADE REPOSITIONING NOT POSSIBLE. BOTH LEADS WERE EXPLANTED. THE HELIX ON THE RV LEAD WOULD NOT FULLY RETRACT SO A NEW RV LEAD WAS USED AND THE PATIENT WAS DOWNGRADED TO A VVI DEVICE DUE TO THE ANATOMICAL CONSTRAINTS SO THERE WAS NO NEED FOR A NEW ATRIAL LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55397 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00072 YR |