FDA Adverse Event Injury Summary report: N

DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

MDR report key: 2952574 · Received February 9, 2013

Report

Report Number
2649622-2013-00623
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 16, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS A COSMETIC DEPRESSION ON THE OUTER INSULATION AND THE OVERLAY TUBING HAD COSMETIC ENVIRONMENTAL STRESS CRACKING. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH IMPEDANCE AND SHORT V-V INTERVALS. THE LEAD WAS PARTIALLY EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57495 DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR