FDA Adverse Event
Malfunction
Summary report: N
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
MDR report key: 2952566
·
Received February 9, 2013
Report
- Report Number
- 2182208-2013-00198
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- December 28, 2012
- Report Date
- December 28, 2012
- Manufacturer
- RICE CREEK MANUFACTURING
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; A PRINTED CIRCUIT BOARD WAS FOUND OUT OF ELECTRICAL SPECIFICATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT WHEN POWERED ON, THE PROGRAMMER FAN COULD BE HEARD RUNNING BUT THAT NOTHING APPEARED ON THE SCREEN. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56864 | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MANUFACTURING | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |