VIRTUOSO DR
Report
- Report Number
- 3004209178-2013-01744
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- April 26, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE RETURNED DEVICE FOUND A HIGH CURRENT DRAIN CONDITION DUE TO CURRENT LEAKAGE IN A CERAMIC CAPACITOR.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076 IMPLANTABLE PACING LEAD, (B)(6) 2002. (B)(4).
IT WAS REPORTED BY THE PATIENT'S SPOUSE THAT THE DEVICE BATTERY DID NOT LAST AS LONG AS EXPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED BY THE PATIENT'S SPOUSE THAT THE DEVICE BATTERY DID NOT LAST AS LONG AS EXPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56209 | VIRTUOSO DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D154AWG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR | Hospitalization| R | 6947 IMPLANTABLE DEFIBRILLATION LEAD |