FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2952514 · Received February 9, 2013

Report

Report Number
2182208-2013-00182
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 21, 2012
Report Date
November 21, 2012
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; THE RF (RADIO FREQUENCY) HEAD CABLE IS FRAYED. IT IS NOTED THE RF HEAD HAD NORMAL TELEMETRY AND PASSED FUNCTIONAL TESTING.

Description of Event or Problem · 1

IT WAS REPORTED THE CORD ON THE PROGRAMMER RF (RADIO-FREQUENCY) HEAD WAS FRAYED. THE RF HEAD WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55635 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2067

Patients

Seq Age Sex Outcome Treatment
1