FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2952307 · Received February 9, 2013

Report

Report Number
2182208-2013-00156
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
December 6, 2012
Report Date
December 17, 2012
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003/S065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; RF (RADIO FREQUENCY) HEAD CABLE FOUND OUT OF ELECTRICAL SPECIFICATION. ANALYSIS ALSO FOUND THE RUBBER PORTION OF THE RF HEAD LABEL MISSING.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER RADIO FREQUENCY HEAD WAS UNABLE TO INTERROGATE. THE HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57219 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2067L

Patients

Seq Age Sex Outcome Treatment
1