FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2952291 · Received February 9, 2013

Report

Report Number
2649622-2013-00475
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 15, 2012
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED ASYSTOLE ASSOCIATED WITH HIGH IMPEDANCE ON THE RIGHT VENTRICULAR (RV) LEAD. IT WAS NOTED THE LEAD WAS FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55823 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R (B)(4) IMPLANTABLE PULSE GENERATOR