FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 2952249 · Received February 9, 2013

Report

Report Number
3004209178-2013-01688
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 15, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY:THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REACHED RECOMMENDED REPLACEMENT TIME (RRT) AND IT WAS EXPECTED TO LAST LONGER. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57731 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154VWC

Patients

Seq Age Sex Outcome Treatment
1 00047 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD