CARELINK PROGRAMMER 2090X
Report
- Report Number
- 2182208-2013-00143
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 21, 2012
- Report Date
- November 21, 2012
- Manufacturer
- RICE CREEK MFG
- Product Code
- NVZ
- PMA / PMN Number
- P890003/S065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE STYLUS/TOUCH SCREEN OUT OF CALIBRATION. AS A RESULT THE OVERLAY/BEZEL ASSEMBLY WAS REPLACED AND THE STYLUS WAS CALIBRATED. (B)(4).
(B)(4). THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE TOUCH PEN FOR THE PROGRAMMER IS OUT OF CALIBRATION, DESPITE HAVING BEEN UPDATED. IT WAS ALSO REPORTED THE X,Y COORDINATES ARE OFF ON THE TOUCH SCREEN. THE STYLUS CALIBRATION WAS RAN AND IT DID NOT HELP. THE PROGRAMMER WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.
IT WAS REPORTED THAT THE TOUCH PEN FOR THE PROGRAMMER IS OUT OF CALIBRATION, DESPITE HAVING BEEN UPDATED. IT WAS ALSO REPORTED THE X,Y COORDINATES ARE OFF ON THE TOUCH SCREEN. THE STYLUS CALIBRATION WAS RAN AND IT DID NOT HELP. THE PROGRAMMER AND RADIOFREQUENCY (RF) HEAD WERE RETURNED FOR REPAIR. BOTH PRODUCTS WERE ANALYZED AND TESTED OUT OF SPECIFICATION. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57566 | CARELINK PROGRAMMER 2090X | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | RICE CREEK MFG | 2090X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |