FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAMMER 2090X

MDR report key: 2952200 · Received February 9, 2013

Report

Report Number
2182208-2013-00143
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 21, 2012
Report Date
November 21, 2012
Manufacturer
RICE CREEK MFG
Product Code
NVZ
PMA / PMN Number
P890003/S065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE STYLUS/TOUCH SCREEN OUT OF CALIBRATION. AS A RESULT THE OVERLAY/BEZEL ASSEMBLY WAS REPLACED AND THE STYLUS WAS CALIBRATED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TOUCH PEN FOR THE PROGRAMMER IS OUT OF CALIBRATION, DESPITE HAVING BEEN UPDATED. IT WAS ALSO REPORTED THE X,Y COORDINATES ARE OFF ON THE TOUCH SCREEN. THE STYLUS CALIBRATION WAS RAN AND IT DID NOT HELP. THE PROGRAMMER WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TOUCH PEN FOR THE PROGRAMMER IS OUT OF CALIBRATION, DESPITE HAVING BEEN UPDATED. IT WAS ALSO REPORTED THE X,Y COORDINATES ARE OFF ON THE TOUCH SCREEN. THE STYLUS CALIBRATION WAS RAN AND IT DID NOT HELP. THE PROGRAMMER AND RADIOFREQUENCY (RF) HEAD WERE RETURNED FOR REPAIR. BOTH PRODUCTS WERE ANALYZED AND TESTED OUT OF SPECIFICATION. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57566 CARELINK PROGRAMMER 2090X PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ RICE CREEK MFG 2090X

Patients

Seq Age Sex Outcome Treatment
1