FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2952176 · Received February 9, 2013

Report

Report Number
3004209178-2013-01675
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 13, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031/S084
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT PRODUCTS: 5072 IMPLANTABLE PACING LEAD (B)(6) 2007; 6947 IMPLANTABLE TACHY LEAD (B)(6) 2009; 4193 IMPLANTABLE PACING LEAD (B)(6) 2009. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) DURING THE WARRANTY PERIOD AND HAD UNEXPECTED LONGEVITY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57526 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00080 YR Hospitalization| R (B)(4) MECHANICAL HEART VALVE