FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX
MDR report key: 2952149
·
Received February 9, 2013
Report
- Report Number
- 2649622-2013-00412
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 13, 2012
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). LEAD REMAINS IN USE.
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT WAS ADMITTED FOR SYNCOPE, AND OVERSENSING, NOISE, AND LOW IMPEDANCE WERE NOTED. POLARITY WAS CHANGED TO UNIPOLAR AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56903 | CAPSUREFIX | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 4068-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | (B)(4) IMPLANTABLE PULSE GENERATOR |