FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2952149 · Received February 9, 2013

Report

Report Number
2649622-2013-00412
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 13, 2012
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). LEAD REMAINS IN USE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS ADMITTED FOR SYNCOPE, AND OVERSENSING, NOISE, AND LOW IMPEDANCE WERE NOTED. POLARITY WAS CHANGED TO UNIPOLAR AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56903 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4068-45

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention (B)(4) IMPLANTABLE PULSE GENERATOR