PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
Report
- Report Number
- 2182208-2013-00135
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- December 5, 2012
- Report Date
- December 13, 2012
- Manufacturer
- RICE CREEK MANUFACTURING
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE PROGRAMMER BOOTED TO A WHITE SCREEN. THE DEVICE HAD AN INTERMITTENT WHITE SCREEN, THE LOW VOLTAGE DIFFERENTIAL SIGNALING (LVDS) WAS LOOSE, THIS WAS RESEATED TO THE SCREEN TO RESOLVE THE ISSUE. IT WAS ALSO NOTED THAT THE MICROPHONE WAS OUT OF SPECIFICATION. PRODUCT ID 2067, SERIAL# (B)(4).
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IT WAS REPORTED THAT THE PROGRAMMER POWERED UP TO A WHITE SCREEN AND THEN IT WOULD NOT DO ANYTHING ELSE. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE PROGRAMMER POWERED UP TO A WHITE SCREEN AND THEN IT WOULD NOT DO ANYTHING ELSE. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55701 | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MANUFACTURING | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) RADIO FREQUENCY PROGRAMMER HEAD |