FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 2952121 · Received February 9, 2013

Report

Report Number
2649622-2013-00395
Event Type
Injury
Date Received
February 9, 2013
Report Date
February 28, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE OUTER INSULATION HAD A DEPRESSION BREACH, THE PROXIMAL CONDUCTOR WAS PULLED/STRETCHED/OVERSTRESSED, THE DISTAL CONDUCTOR WAS PULLED/STRETCHED/OVERSTRESSED, THE PROXIMAL CONDUCTOR WAS MELTED, THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR, THERE WAS BLOOD ON THE DISTAL CONDUCTOR, THE OUTER INSULATION WAS MELTED, AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION; FULL LEAD IN SEGMENTS RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND HIGH THRESHOLD, WITH A LEAD WARNING AND POLARITY SWITCH. THE RV (RIGHT VENTRICULAR) LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56707 CAPSURE SP ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5024M-58

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Hospitalization| R