CAPSURE SP
Report
- Report Number
- 2649622-2013-00395
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- February 28, 2013
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
EVALUATION SUMMARY: THE OUTER INSULATION HAD A DEPRESSION BREACH, THE PROXIMAL CONDUCTOR WAS PULLED/STRETCHED/OVERSTRESSED, THE DISTAL CONDUCTOR WAS PULLED/STRETCHED/OVERSTRESSED, THE PROXIMAL CONDUCTOR WAS MELTED, THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR, THERE WAS BLOOD ON THE DISTAL CONDUCTOR, THE OUTER INSULATION WAS MELTED, AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION; FULL LEAD IN SEGMENTS RETURNED AND ANALYZED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND HIGH THRESHOLD, WITH A LEAD WARNING AND POLARITY SWITCH. THE RV (RIGHT VENTRICULAR) LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56707 | CAPSURE SP | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5024M-58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR | Hospitalization| R |