VIRTUOSO DR
Report
- Report Number
- 3004209178-2013-01665
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- June 11, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA FROM THE DEVICE WAS RECEIVED AND REVIEWED. NO ANOMALIES WERE FOUND. CONCOMITANT DEVICE: 5076 IMPLANTABLE PACING LEAD: (B)(6) 2008. (B)(4).
IT WAS REPORTED THAT THE DEFIBRILLATOR WAS REMOVED AND REPLACED DUE TO PREMATURE BATTERY DEPLETION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56119 | VIRTUOSO DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D154AWG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00066 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD |