CONCERTO II CRT-D
Report
- Report Number
- 3004209178-2013-01656
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 7, 2012
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: (B)(4). THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 80% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. THE PRIMARY SAVE TO DISK ANALYSIS NOTED LONGEVITY AND BATTERY DATA DEPLETION INDICATED ELECTIVE REPLACEMENT INDICATOR (ERI). CONCOMITANT CONTINUED: 4193 IMPLANTABLE PACING LEAD: (B)(6) 2007. 4076 IMPLANTABLE PACING LEAD: (B)(6) 2007. (B)(4).
IT WAS REPORTED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE CUSTOMER QUESTIONED THE LONGEVITY EXPECTATION OF THE DEVICE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55772 | CONCERTO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D274TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | 6949 IMPLANTABLE TACHY LEAD |