CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2013-00419
- Event Type
- Injury
- Date Received
- February 9, 2013
- Date of Event
- November 12, 2012
- Report Date
- November 13, 2012
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD ON THE DISTAL CONDUCTOR, THE OUTER INSULATION WAS BREACHED CUT, THE HELIX WAS DISTORTED/BENT, THE HELIX APPEARS TO HAVE BEEN PULLED/STRETCHED/OVERSTRESSED AND THERE WAS APPARENT EXPLANT DAMAGE. CONCOMITANT MEDICAL PRODUCTS: RVDR01 IMPLANTABLE PULSE GENERATOR: (B)(6) 2012. (B)(4).
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD DISLODGED SHORTLY AFTER IMPLANT. THE DISLODGEMENT WAS FOUND ON X-RAY THE NEXT MORNING. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55661 | CAPSUREFIX NOVUS MRI | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR | Hospitalization| R | 5086MRI IMPLANTABLE PACING LEAD |