FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2952029 · Received February 9, 2013

Report

Report Number
2649622-2013-00419
Event Type
Injury
Date Received
February 9, 2013
Date of Event
November 12, 2012
Report Date
November 13, 2012
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD ON THE DISTAL CONDUCTOR, THE OUTER INSULATION WAS BREACHED CUT, THE HELIX WAS DISTORTED/BENT, THE HELIX APPEARS TO HAVE BEEN PULLED/STRETCHED/OVERSTRESSED AND THERE WAS APPARENT EXPLANT DAMAGE. CONCOMITANT MEDICAL PRODUCTS: RVDR01 IMPLANTABLE PULSE GENERATOR: (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD DISLODGED SHORTLY AFTER IMPLANT. THE DISLODGEMENT WAS FOUND ON X-RAY THE NEXT MORNING. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55661 CAPSUREFIX NOVUS MRI DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Hospitalization| R 5086MRI IMPLANTABLE PACING LEAD