FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX
MDR report key: 2951922
·
Received February 9, 2013
Report
- Report Number
- 2649622-2013-00294
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 8, 2012
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039/S2
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5068, IMPLANTABLE PACING LEAD, (B)(6) 1999. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD DEVELOPED HIGH THRESHOLDS OVER TIME AND APPEARED TO HAVE SIGNS OF CLAVICLE CRUSH. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56099 | CAPSUREFIX | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5068-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00030 YR | ADDR01 IMPLANTABLE PULSE GENERATOR |