MDR report key: 2951898 · Received February 9, 2013

Report

Report Number
2649622-2013-00264
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 6, 2012
Report Date
November 6, 2012
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015/S012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE DISTAL CONDUCTOR WAS KINKED/BUCKLED. (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE GUIDEWIRE COULD NOT BE REMOVED FROM THE LEFT VENTRICULAR LEAD, NOR COULD THE GUIDEWIRE BE ADVANCED. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56756 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419488

Patients

Seq Age Sex Outcome Treatment
1 00079 YR 6949 IMPLANTABLE TACHY LEAD