FDA Adverse Event
Malfunction
Summary report: N
ATTAIN OTW
MDR report key: 2951898
·
Received February 9, 2013
Report
- Report Number
- 2649622-2013-00264
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 6, 2012
- Report Date
- November 6, 2012
- Manufacturer
- MPRI
- Product Code
- LWP
- PMA / PMN Number
- P010015/S012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE DISTAL CONDUCTOR WAS KINKED/BUCKLED. (B)(6). (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE GUIDEWIRE COULD NOT BE REMOVED FROM THE LEFT VENTRICULAR LEAD, NOR COULD THE GUIDEWIRE BE ADVANCED. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56756 | ATTAIN OTW | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MPRI | 419488 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00079 YR | 6949 IMPLANTABLE TACHY LEAD |