FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2951883 · Received February 9, 2013

Report

Report Number
2649622-2013-00256
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 6, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED AN INAPPROPRIATE SHOCK FROM THE RIGHT VENTRICULAR (RV) LEAD DUE TO T-WAVE OVERSENSING. THE POLARITY WAS REPROGRAMMED AND THE RV LEAD IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56757 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR