FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2951871 · Received February 9, 2013

Report

Report Number
2649622-2013-00278
Event Type
Injury
Date Received
February 9, 2013
Date of Event
April 8, 2012
Report Date
February 22, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039/S017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, APPARENT EXPLANT DAMAGE WAS NOTED. (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, APPARENT EXPLANT DAMAGE WAS NOTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 6947 IMPLANTABLE DEFIBRILLATING LEAD 2012 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BOTH THE RIGHT VENTRICULAR AND THE ATRIAL LEADS DISLODGED. BOTH LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56684 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407645

Patients

Seq Age Sex Outcome Treatment
1 00033 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (IC