FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 2951835 · Received February 9, 2013

Report

Report Number
3004209178-2013-01617
Event Type
Injury
Date Received
February 9, 2013
Report Date
February 1, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). CONCOMITANT PRODUCTS: 6944 IMPLANTABLE TACHY LEAD, (B)(6) 2008.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS AT ELECTIVE REPLACEMENT INDICATOR AND HAD PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55558 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00042 YR Hospitalization| R 4076 IMPLANTABLE PACING LEAD