CONCERTO CRT-D DR
Report
- Report Number
- 3004209178-2013-01613
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 5, 2012
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4193 IMPLANTABLE PACING LEAD (B)(6) 2004; 6947 IMPLANTABLE TACHY LEAD (B)(6) 2004. (B)(4).
PRODUCT EVENT SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT HAD MULTIPLE SHOCKS FROM A POSSIBLE RIGHT VENTRICULAR (RV) LEAD FRACTURE. NOISE AND OVERSENSING WERE NOTED ON THE RV LEAD. EARLY ELECTIVE REPLACEMENT INDICATOR OF THE DEVICE WAS ALSO REPORTED POSSIBLY DUE TO THE MULTIPLE SHOCKS. THE RV LEAD AND DEVICE WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57587 | CONCERTO CRT-D DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | C154DWK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00068 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD |