FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2951721 · Received February 9, 2013

Report

Report Number
2649622-2013-00190
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 6, 2012
Report Date
November 6, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015/S24
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED; ANALYSIS REVEALED THE HELIX WAS DISTORTED/BENT. THERE WAS BLOOD ON THE DISTAL ELECTRODE AND THE LEAD WAS DAMAGED AT IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE RIGHT VENTRICULAR LEAD HELIX DID NOT APPEAR TO BE EXTENDED UNDER FLUOROSCOPY. THE LEAD WAS REMOVED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57276 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00078 YR 5076 IMPLANTABLE PACING LEAD