FDA Adverse Event
Injury
Summary report: N
VIRTUOSO DR
MDR report key: 2951707
·
Received February 9, 2013
Report
- Report Number
- 3004209178-2013-01581
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 5, 2012
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. (B)(4). CONCOMITANT MEDICAL PRODUCTS: IMPLANTABLE TACHY LEAD: (B)(6) 1998.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE LONGEVITY WAS LESS THAN FIVE YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56385 | VIRTUOSO DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D154AWG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00066 YR | Hospitalization| R | 6945 IMPLANTABLE TACHY LEAD |