FDA Adverse Event Injury Summary report: N

DA+ C SERIES DR

MDR report key: 2951631 · Received February 9, 2013

Report

Report Number
9614453-2013-00050
Event Type
Injury
Date Received
February 9, 2013
Report Date
December 2, 2012
Manufacturer
IPG MFG SWITZERLAND
Product Code
DXY
PMA / PMN Number
P990001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) PREMATURELY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55710 DA+ C SERIES DR PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY IPG MFG SWITZERLAND C50A1

Patients

Seq Age Sex Outcome Treatment
1 00069 YR Hospitalization| R