FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 2951508 · Received February 9, 2013

Report

Report Number
3004209178-2013-01551
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 20, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT IS BEING REPORTED BASED ON ANALYSIS RESULTS. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCTS: 4076, IMPLANTABLE PACING LEAD, - (B)(6) 2007. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS EXPLANTED AND NOT REPLACED PER THE PATIENT'S REQUEST. THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56601 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00030 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD