Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

FURTHER REVIEW PROMPTED A CHANGE IN THE DEVICE ANALYSIS RESULTS.

Additional Manufacturer Narrative · 1

(B)(4). THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE PROGRAMMER WAS RETURNED AND ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT IT WOULD NOT BOOT COMPLETELY, THERE WAS DEVICE LOCK-UP AND THE PRINTED CIRCUIT BOARD WAS OUT OF SPECIFICATION. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER FAILED TO BOOT UP. THE LOGO WAS DISPLAYED, BUT DID NOT CONTINUE WITH THE START-UP. TECHNICAL SERVICES (TS) RECOMMENDED THAT THE SERVICE DISK BE TRIED. THIS CLIENT TRIED THE DISK AND THE PROGRAMMER STILL "HANGS" AND THE SERVICE DISK DID NOT RUN. THE PROGRAMMER WILL BE RETURNED FOR SERVICE. ANOTHER PROGRAMMER WAS USED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

THE PROGRAMMER WAS RETURNED FOR SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER FAILED TO BOOT UP. THE LOGO WAS DISPLAYED, BUT DID NOT CONTINUE WITH THE START-UP. TECHNICAL SERVICES (TS) RECOMMENDED THAT THE SERVICE DISK BE TRIED. THIS CLIENT TRIED THE DISK AND THE PROGRAMMER STILL "HANGS" AND THE SERVICE DISK DID NOT RUN. THE PROGRAMMER WILL BE RETURNED FOR SERVICE. ANOTHER PROGRAMMER WAS USED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.