FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 2951450 · Received February 9, 2013

Report

Report Number
2182208-2013-00037
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 2, 2012
Report Date
November 2, 2012
Manufacturer
RICE CREEK MANUFACTURING
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE PROGRAMMER HAD A BLANK SCREEN, THE FLEX TAPE WAS CREASED CAUSING IT TO CRACK AND FAIL. IT WAS ALSO NOTED THAT THE DISPLAY HASPS WERE BROKEN AND THE RIGHT ANTENNA WAS OUT OF SPECIFICATION. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER POWERED UP TO A BLACK SCREEN. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER POWERED UP TO A BLACK SCREEN. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55891 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MANUFACTURING 2090W

Patients

Seq Age Sex Outcome Treatment
1