OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-02557
- Event Type
- Malfunction
- Date Received
- February 8, 2013
- Report Date
- January 17, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 2 (08/14/2013)-PRODUCT EVALUATION: THE ANALYSIS OF THE SUBJECT METER WAS COMPLETED ON (B)(4) 2013 BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE REPORTED COMPLAINT COULD NOT BE REPRODUCED. THE METER MET SPECIFICATIONS FOR TESTING. NO ISSUES WERE IDENTIFIED DURING INVESTIGATION.
THE SUBJECT METER AND TEST STRIPS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION. THE EVALUATION OF THE PRODUCTS HAVE NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
(B)(4): THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. A SECONDARY ISSUE UNRELATED TO THE COMPLAINT WAS FOUND, THERE WERE EXTRA TEST STRIPS IN THE TEST STRIP VIAL. THE METER WAS ALSO RETURNED ON (B)(4) 2013 HOWEVER EVALUATION OF THE PRODUCT HAS NOT YET BEEN COMPLETED, THEREFORE NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, ANOTHER FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATE RESULTS WITH CONTROL SOLUTION, WITH A READING OF "9.4MMOL/L." THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54086 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3404895 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |