FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2951222 · Received February 8, 2013

Report

Report Number
3008382007-2013-02557
Event Type
Malfunction
Date Received
February 8, 2013
Report Date
January 17, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2 (08/14/2013)-PRODUCT EVALUATION: THE ANALYSIS OF THE SUBJECT METER WAS COMPLETED ON (B)(4) 2013 BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE REPORTED COMPLAINT COULD NOT BE REPRODUCED. THE METER MET SPECIFICATIONS FOR TESTING. NO ISSUES WERE IDENTIFIED DURING INVESTIGATION.

Additional Manufacturer Narrative · 1

THE SUBJECT METER AND TEST STRIPS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION. THE EVALUATION OF THE PRODUCTS HAVE NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

(B)(4): THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. A SECONDARY ISSUE UNRELATED TO THE COMPLAINT WAS FOUND, THERE WERE EXTRA TEST STRIPS IN THE TEST STRIP VIAL. THE METER WAS ALSO RETURNED ON (B)(4) 2013 HOWEVER EVALUATION OF THE PRODUCT HAS NOT YET BEEN COMPLETED, THEREFORE NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, ANOTHER FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATE RESULTS WITH CONTROL SOLUTION, WITH A READING OF "9.4MMOL/L." THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54086 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3404895

Patients

Seq Age Sex Outcome Treatment
1 60 YR