ENDURANT II ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2013-00233
- Event Type
- Injury
- Date Received
- February 8, 2013
- Date of Event
- January 16, 2013
- Report Date
- March 21, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). RESULTS: ENDOLEAK. ANGULATED AORTIC NECK. STENT GRAFT WAS USED FOR TREATMENT OF A PATIENT WITH AN AORTIC NECK ANGULATION OF 100 DEGREES. CONCLUSION: ANGULATED SHAPED AORTIC NECK. STENT GRAFT WAS USED FOR TREATMENT OF A PATIENT WITH AN AORTIC NECK ANGULATION OF 100 DEGREES.
METHOD: (FILMS).
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY THREE WEEKS AGO. THE AORTIC NECK ANGLE HAD AN ANGULATION OF 100 DEGREES AND IT WAS 30MM LONG. IT WAS REPORTED THAT INTRA-OPERATIVELY A TYPE I ENDOLEAK WAS OBSERVED. THE STENT GRAFT WAS BALLOONED WITH A RELIANT BALLOON, AS WELL AS WITH 26MM AND 28MM BALLOONS FROM ANOTHER MANUFACTURER. DUE TO THE SUPRA RENAL NECK ANGLE IT WAS NOT POSSIBLE TO FULLY INFLATE THE OTHER MANUFACTURER'S BALLOONS AND THE ENDOLEAK WAS STILL PRESENT. THE DECISION WAS MADE TO IMPLANT A PALMAZ STENT. THIS REDUCED THE ENDOLEAK BUT DID NOT FULLY RESOLVE IT AND THE DECISION WAS MADE TO MONITOR THE PATIENT. A FOLLOW-UP CT SCAN WAS DONE A FEW DAYS LATER AND SHOWED THAT THE ENDOLEAK HAD RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
FILMS WERE RECEIVED AND THEIR EVALUATION WAS COMPLETED. SEVERAL STILL ANGIOGRAM IMAGES DURING IMPLANT WERE REVIEWED. THE PROXIMAL AORTIC NECK IS SEVERELY ANGULATED IN THE VIEW PROVIDED; THE NECK APPEARS OFF-LABEL. THE TIP OF THE DELIVERY SYSTEM IS BIASED AGAINST THE RIGHT LATERAL WALL OF THE SUPRARENAL AORTA. POST-IMPLANT IMAGE SHOWS THAT THE BIFURCATE WAS PLACED JUST BELOW THE RENAL ARTERIES IN THE ANGULATED NECK. THERE IS CONTRAST SEEN IN THE ANEURYSM SAC. THE TYPE OF ENDOLEAK CANNOT BE DETERMINED FROM THE IMAGES PROVIDED. NO OTHER STENT GRAFT ISSUES ARE SEEN. THE FINAL IMAGE POST-IMPLANT WITH THE PALMAZ PLACED INTO THE AORTIC BODY SHOWS A POSSIBLE REDUCTION OF THE AMOUNT OF CONTRAST SEEN IN THE ANEURYSM SAC. THE CAUSE OF THE REPORTED PROXIMAL TYPE I ENDOLEAK COULD NOT BE DETERMINED; HOWEVER, IT IS LIKELY THAT THE SEVERELY ANGULATED PROXIMAL NECK CONTRIBUTING FACTOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54636 | ENDURANT II ABDOMINAL STENT GRAFT SYSTEM | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V03013640 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |