COBE SPECTRA BLOOD COLLECTION
Report
- Report Number
- 1722028-2013-00048
- Date Received
- February 8, 2013
- Date of Event
- December 14, 2012
- Report Date
- January 10, 2013
- Manufacturer
- TERUMO BCT
- Product Code
- GKT
- PMA / PMN Number
- BK020041
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION: A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THIS UNIT SHOWED NO IRREGULARITIES DURING MANUFACTURING THAT WERE RELEVANT TO THIS ISSUE. THE SYMPTOMS AND SEVERITY OF THE PATIENT REACTION WERE INCONSISTENT WITH REACTIONS TO CITRATE AND ETHYLENE OXIDE. THERE HAVE ALSO BEEN NO OTHER REPORTS OF SIMILAR PATIENT REACTIONS ON THIS LOT OF DISPOSABLES, OTHER THAN THE ONES REPORTED BY THIS CUSTOMER SITE. THE DISPOSABLE SET WAS UNAVAILABLE FOR RETURN AND INVESTIGATION. THE CUSTOMER WAS USING CALCIUM GLUCONATE DURING THE PROCEDURE. THIS SOLUTION CONSISTED OF TWO BLISTERS THAT WERE DILUTED WITH 250CC OF SALINE AND WAS ADMINISTERED DIRECTLY THROUGH THE RETURN LINE. THIS IS PER THE CUSTOMER'S NORMAL PROCEDURE. AFTER THE PROCEDURE, IT WAS FOUND THAT THE CALCIUM GLUCONATE BLISTERS THAT WERE BEING USED DURING THIS PROCEDURE HAD BEEN RECALLED BY THE REGULATORY BODY IN ARGENTINA DUE TO MICROBIAL CONTAMINATION. ROOT CAUSE: OPERATOR ERROR OF USING RECALLED CALCIUM GLUCONATE. THE CALCIUM GLUCONATE THAT WAS USED DURING THE PROCEDURE HAD MICROBIAL CONTAMINATION.
THE CUSTOMER REPORTED THAT THEIR PATIENT HAD TREMORS, FEVER, TACHYPNEA AND TACHYCARDIA DURING THE PROCEDURE. THE PATIENT REQUIRED MEDICAL INTERVENTION IN THE FORM OF STEROIDS. ALL THE BLOOD CULTURE AND PRODUCTS CULTURE TAKEN WERE NEGATIVE. THE DISPOSABLE SET IS NOT AVAILABLE FOR EVALUATION BECAUSE THE CUSTOMER DISCARDED IT. THIS REPORT IS BEING FILED DUE TO MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54582 | COBE SPECTRA BLOOD COLLECTION | COBE SPECTRA WHITE BLOOD CELL SET, CLOSED | GKT | TERUMO BCT | 02U15275 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00047 YR | Other |