(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. THIS DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A DAMAGED BATTERY WAS CONFIRMED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE WAS IDENTIFED AS A DEFECTIVE MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO RESOLVE THE REPORTE CONDITION. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.
(B)(4).
THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A DAMAGED BATTERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.