UV FLASH TRANSFER SETS
Report
- Report Number
- 1416980-2013-03359
- Event Type
- Malfunction
- Date Received
- February 8, 2013
- Date of Event
- January 24, 2013
- Report Date
- January 24, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
(B)(4). THIS IS A PRODUCT NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510K NUMBER. THE SAMPLE HAS BEEN REPORTED TO BE AVAILABLE FOR EVALUATION AND HAS BEEN REQUESTED. HOWEVER, THE SAMPLE HAS NOT BEEN RECEIVED FOR EVALUATION AS OF YET. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THIS COMPLAINT WAS CONFIRMED DURING EVALUATION OF THE RETURNED SAMPLE; HOWEVER, NO ROOT CAUSE COULD BE DETERMINED. A VISUAL INSPECTION WAS PERFORMED AND A CUT IN THE TUBING WAS NOTED. LEAK TESTING WAS PERFORMED AND A CUT IN THE TUBING WAS NOTED. A CLEAR PASSAGE TEST AND CLAMP FUNCTION TEST WERE PERFORMED WITH NO ISSUES NOTED. A BATCH REVIEW COULD NOT BE PERFORMED BECAUSE THE LOT NUMBER WAS NOT AVAILABLE.
A NURSE REPORTED TO BAXTER JAPAN THAT DURING USE OF A TRANSFER SET, THERE WAS A LEAK NOTICED FROM THE SILICONE TUBING OF THE TRANSFER SET. THE NURSE STATED THAT THERE WAS A PIN-SIZED HOLE IN THE SILICONE TUBING, AND THAT IS FROM WHERE THE LEAK WAS COMING. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54239 | UV FLASH TRANSFER SETS | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |