FDA Adverse Event Malfunction Summary report: N

UV FLASH TRANSFER SETS

MDR report key: 2950910 · Received February 8, 2013

Report

Report Number
1416980-2013-03359
Event Type
Malfunction
Date Received
February 8, 2013
Date of Event
January 24, 2013
Report Date
January 24, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A PRODUCT NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510K NUMBER. THE SAMPLE HAS BEEN REPORTED TO BE AVAILABLE FOR EVALUATION AND HAS BEEN REQUESTED. HOWEVER, THE SAMPLE HAS NOT BEEN RECEIVED FOR EVALUATION AS OF YET. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT WAS CONFIRMED DURING EVALUATION OF THE RETURNED SAMPLE; HOWEVER, NO ROOT CAUSE COULD BE DETERMINED. A VISUAL INSPECTION WAS PERFORMED AND A CUT IN THE TUBING WAS NOTED. LEAK TESTING WAS PERFORMED AND A CUT IN THE TUBING WAS NOTED. A CLEAR PASSAGE TEST AND CLAMP FUNCTION TEST WERE PERFORMED WITH NO ISSUES NOTED. A BATCH REVIEW COULD NOT BE PERFORMED BECAUSE THE LOT NUMBER WAS NOT AVAILABLE.

Description of Event or Problem · 1

A NURSE REPORTED TO BAXTER JAPAN THAT DURING USE OF A TRANSFER SET, THERE WAS A LEAK NOTICED FROM THE SILICONE TUBING OF THE TRANSFER SET. THE NURSE STATED THAT THERE WAS A PIN-SIZED HOLE IN THE SILICONE TUBING, AND THAT IS FROM WHERE THE LEAK WAS COMING. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54239 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1