CORE IMPACTION DRILL

MDR report key: 2950756 · Received February 8, 2013

Report

Report Number
0001811755-2013-00211
Event Type
Malfunction
Date Received
February 8, 2013
Date of Event
January 22, 2013
Report Date
January 22, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, CORROSION WAS DISCOVERED THROUGHOUT THE INTERNAL COMPONENTS OF THE DEVICE, INCLUDING THE MOTOR, DRIVESHAFT ASSEMBLY AND ROTOR. INCREASED FRICTION DUE TO CORROSION IS A PROBABLE CAUSE OF THE REPORTED OVERHEATING.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.