FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2950756
·
Received February 8, 2013
Report
- Report Number
- 0001811755-2013-00211
- Event Type
- Malfunction
- Date Received
- February 8, 2013
- Date of Event
- January 22, 2013
- Report Date
- January 22, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVALUATION, CORROSION WAS DISCOVERED THROUGHOUT THE INTERNAL COMPONENTS OF THE DEVICE, INCLUDING THE MOTOR, DRIVESHAFT ASSEMBLY AND ROTOR. INCREASED FRICTION DUE TO CORROSION IS A PROBABLE CAUSE OF THE REPORTED OVERHEATING.
Description of Event or Problem · 1
IT WAS REPORTED THAT AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54774 | CORE IMPACTION DRILL | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |