XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2013-00751
- Event Type
- Injury
- Date Received
- February 8, 2013
- Date of Event
- December 31, 2012
- Report Date
- January 16, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P070015
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THERE WAS NO REPORTED PRODUCT DEFICIENCY AND THE PRODUCT WAS NOT RETURNED. THE REPORTED PATIENT EFFECT OF DISSECTION, AS LISTED IN THE XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM INSTRUCTIONS FOR USE (IFU) IS A KNOWN ADVERSE EVENT ASSOCIATED WITH CORONARY STENTING PROCEDURES. ALTHOUGH A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECTS AND THE RELATIONSHIP TO THE DEVICE, IF ANY, CANNOT BE DETERMINED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.
IT WAS REPORTED THAT DURING A PROCEDURE OF THE RIGHT CORONARY ARTERY (RCA), THE 3.0 X 15 MM XIENCE V STENT WAS DEPLOYED, HOWEVER, A DISSECTION WAS NOTED. A SECOND XIENCE STENT WAS PLACED AS TREATMENT. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54560 | XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 2070441 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |