FLOGARD
Report
- Report Number
- 1416980-2013-03258
- Event Type
- Malfunction
- Date Received
- February 8, 2013
- Date of Event
- January 1, 2013
- Report Date
- January 15, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH A LOW BATTERY ALARM WAS CONFIRMED BY BAXTER SERVICE PERSONNEL. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE A DEFECTIVE MAIN BATTERY. THE BATTERY WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP THAT PRESENTED A LOW BATTERY ALARM. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.