FDA Adverse Event Injury Summary report: N

MINICAP TRANSFER SET

MDR report key: 2950239 · Received February 8, 2013

Report

Report Number
1416980-2013-03255
Event Type
Injury
Date Received
February 8, 2013
Date of Event
January 1, 2013
Report Date
January 18, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: (B)(4). THIS IS REPORT 2 OF 2 INVOLVED IN THIS PERITONITIS INCIDENT. THIS REPORT OF PERITONITIS WAS RECEIVED WITH NO ALLEGED DEVICE MALFUNCTION OR USE ERROR; THEREFORE, A SAMPLE WAS NOT REQUESTED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A BATCH REVIEW WAS CONDUCTED FOR THE POTENTIALLY ASSOCIATED LOT NUMBER H12G09010, H12H31095, H12H07087, H12J15038, H12J01053, AND H12J11037. NO EXCEPTIONS WERE OBSERVED THAT WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE PROBLEM WAS NOT CONFIRMED AND THE CAUSE OF THE PERITONITIS WAS UNDETERMINED. NO DEVICE MALFUNCTION OR USE ERROR WAS IDENTIFIED.

Description of Event or Problem · 1

THIS REPORT WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE (GPV) AND IS A SPONTANEOUS REPORT BY A CONSUMER WITH SUPPLEMENTAL INFORMATION PROVIDED BY THE NURSE IN THE USA OF PERITONITIS IN A PATIENT COINCIDENT WITH DIANEAL THERAPIES FOR PERITONEAL DIALYSIS (PD). DIANEAL THERAPIES WERE ONGOING. DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING WAS REPORTED. ON AN UNREPORTED DATE, THE PATIENT EXPERIENCED CONSTIPATION WHICH LED TO PERITONITIS ON 12JAN2013. THE PATIENT WAS NOT HOSPITALIZED FOR THE EVENT. ON AN UNREPORTED DATE, THE PATIENT WAS TREATED WITH TWO ANTIBIOTICS, INTRA-PERITONEALLY (SPECIFICS UNKNOWN). THE CAUSE OF THE PERITONITIS WAS UNKNOWN BUT REPORTED TO NOT BE DUE TO A BREAK IN ASEPTIC TECHNIQUE. THE PATIENT WAS RECOVERING FROM THIS PERITONITIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54483 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention DIANEAL PD4 AMBUFLEX| DIANEAL PD4 ULTRABAG| HOMECHOICE