FLOGARD
Report
- Report Number
- 1416980-2013-03247
- Event Type
- Malfunction
- Date Received
- February 8, 2013
- Date of Event
- January 1, 2013
- Report Date
- January 11, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH AN F-38 ALARM WAS CONFIRMED AND REPAIRED BY BAXTER SERVICE PERSONNEL ONSITE AT THE CUSTOMER FACILITY. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE DAMAGED FORCE-SENSING RESISTORS. THE FORCE-SENSING RESISTORS WERE REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP THAT PRESENTED A "BATTERY F-38 ALARM". IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.