FDA Adverse Event Injury Summary report: N

RESOLUTE INTEGRITY RX

MDR report key: 2950015 · Received February 8, 2013

Report

Report Number
9612164-2013-00146
Event Type
Injury
Date Received
February 8, 2013
Date of Event
January 10, 2013
Report Date
January 11, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
NIQ
PMA / PMN Number
P110013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION RESULTS AND CONCLUSION: DISSECTION. SEVERE TORTUOSITY AND MODERATE CALCIFICATION EVALUATION. (B)(4).

Description of Event or Problem · 1

A RESOLUTE INTEGRITY DRUG ELUTING STENT (3.5X18MM) WAS DEPLOYED IN PROXIMAL RCA WHICH WAS REPORTED TO EXHIBIT SEVERE TORTUOSITY, MODERATE CALCIFICATION AND 30% STENOSIS FOLLOWING PRE- DILATATION. POST DEPLOYMENT A DISTAL EDGE DISSECTION WAS NOTICED .THE PHYSICIAN ATTEMPTED TO IMPLANT A SECOND RESOLUTE INTEGRITY DRUG ELUTING STENT (2.25X8MM) IN TREATMENT . NO ABNORMALITY WAS NOTED DURING PREPARATION OF THE RESOLUTE DEVICE; HOWEVER, DURING ATTEMPTED DELIVERY THE RESOLUTE DISLODGED PRIOR TO THE LESION IN THE OSTIUM; PROXIMAL TO PREVIOUSLY DEPLOYED STENT. A SPRINTER LEGEND BALLOON WAS USED TO GET INSIDE THE DISLODGED RESOLUTE. THIS WAS THEN PRECEDED WITH MULTIPLE BALLOON SIZES AND WENT ALL THE WAY UP TO A 3.5 BALLOON TO EXPAND STENT. PATIENT IS REPORTED TO BE FINE POST PROCEDURE. PROCEDURAL IMAGE REVIEW: THE PROCEDURAL IMAGES SHOW A PROXIMAL RCA LESION WITH SIGNIFICANT STENOSIS AND TORTUOSITY. PRE-DILATIONS ARE PERFORMED AND A STENT IS DEPLOYED. A DISSECTION CAN BE SEEN AT THE DISTAL EDGE OF THE DEPLOYED STENT. THE FOLLOWING IMAGES SHOW A DISLODGED STENT PROXIMAL TO THE DEPLOYED STENT AT THE OSTIUM OF THE RCA. THIS DISLODGED STENT IS SUCCESSFULLY POSITIONED AND DEPLOYED DISTAL TO THE DEPLOYED STENT TO TREAT THE DISSECTION. THE FINAL CAG SHOWS A GOOD OUTCOME

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54291 RESOLUTE INTEGRITY RX STENT, CORONARY, DRUG-ELUTING NIQ MEDTRONIC IRELAND

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Required Intervention