ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00082
- Event Type
- Malfunction
- Date Received
- February 7, 2013
- Date of Event
- January 8, 2013
- Report Date
- January 10, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4) RETURNED BREATHING CIRCUITS: CIRCUITS 1-4, LOT: 120731; CIRCUITS 5-7, LOT: UNKNOWN. METHOD: SEVEN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WERE VISUALLY INSPECTED FOR THE REPORTED DAMAGE. THE INSPIRATORY AND EXPIRATORY LIMBS OF CIRCUIT 7 WERE CHECKED TO SEE IF THEY COULD BE CONNECTED TO A HEATER WIRE ADAPTOR. RESULTS: VISUAL INSPECTION REVEALED A HOLE IN THE EXPIRATORY LIMB IN THE FOLLOWING CIRCUITS: CIRCUIT 1: APPROXIMATELY 17CM FROM THE PROXIMAL CONNECTOR; CIRCUIT 2: APPROXIMATELY 27CM FROM THE PROXIMAL CONNECTOR; CIRCUIT 3: APPROXIMATELY 38CM FROM THE PROXIMAL CONNECTOR; CIRCUIT 4: APPROXIMATELY 48CM FROM THE PROXIMAL CONNECTOR; CIRCUIT 5: APPROXIMATELY 5.5CM FROM THE PROXIMAL CONNECTOR; CIRCUIT 6: APPROXIMATELY 11CM FROM THE PROXIMAL CONNECTOR; CIRCUIT 7: APPROXIMATELY 14CM FROM THE PROXIMAL CONNECTOR. THE HEATER WIRE ADAPTOR COULD NOT BE FULLY INSERTED TO THE INSPIRATORY AND EXPIRATORY LIMB HEATER WIRE PINS OF CIRCUIT 7. BOTH THE INSPIRATORY AND EXPIRATORY HEATER WIRE PINS WERE BENT AND OUT OF SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120731. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED PRIOR TO LEAVING THE PRODUCTION FACILITY. ANY BREATHING CIRCUITS THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE LEAKS ON THE RETURNED RT340 BREATHING CIRCUITS DEVELOPED POST PRODUCTION. THE IDENTIFIED HOLES IN THE EXPIRATORY LIMBS WERE MOST LIKELY CAUSED BY A SCRATCH OR PUNCTURE BY A BLUNT OBJECT. THE KEY DIFFERENCE BETWEEN THE EVAQUA BREATHING CIRCUITS AND CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY TUBE OF EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOUR FROM EXPIRED VENTILATORY GAS TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER THE EVAQUA TUBING REMAINS MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY NON-FPH CIRCUIT HANGERS. ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. IT IS POSSIBLE FOR THE USER TO BEND OR SPLIT THE HEATER WIRE PINS IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR BENT OR SPLIT PINS REPORTED TO US BY HEALTHCARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE BENT OR SPLIT DURING PRODUCTION OR BY THE END USER. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.
A HOSPITAL IN (B)(6) REPORTED TO OUR DISTRIBUTOR THAT SEVEN RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUITS HAD HOLES IN THE EXPIRATORY LIMB. THEY FURTHER REPORTED THAT ONE OF THE CIRCUITS HAD BENT HEATER WIRE PINS. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51888 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 120731 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |