FDA Adverse Event Injury Summary report: N

ARTICULEZE M HEAD 36MM +1.5

MDR report key: 2948347 · Received February 7, 2013

Report

Report Number
1818910-2013-12006
Event Type
Injury
Date Received
February 7, 2013
Date of Event
January 7, 2013
Report Date
January 9, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
PK980513
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION HAS BEEN REOPENED DUE TO RECEIVING PART AND LOT NUMBER. DEPUY WILL NOTIFY THE FDA WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE AND/OR DHR REVIEW WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL RELATED REPORTS FOR THE LOT CODES 2213913, A3PD21000, AF2FF4000, ZL9LT4000, X5FEC1000, Y2REG4000, Z5NKW4000, AND X52CX4000. A SEARCH OF THE COMPLAINT DATABASE FINDS ADDITIONAL REPORTED INCIDENTS AGAINST LOT CODE AA5DA4000 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. A SEARCH OF THE COMPLAINT DATABASE FINDS ADDITIONAL REPORTED INCIDENTS AGAINST LOT CODE 1901192 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, INFLAMMATION, DIFFICULTY AMBULATING, AND DISABILITY. IT IS ALSO ALLEGED THAT THE PATIENT FELT THE HIP MOVE OUT OF PLACE. AN X-RAY CONFIRMED THAT THE ACETABULAR COMPONENT MOVED. **UPDATE: 4/25/2013 PFS WAS RECEIVED FROM LEGAL, MEDICAL RECORDS WERE RECEIVED FROM LEGAL, AND PART/LOT INFORMATION WAS IDENTIFIED. RECORDS INDICATE THAT THE REVISION ON (B)(6) 2006 WAS DUE TO A LOOSE ACETABULAR SHELL. THE REVISION ON (B)(6) 2007 WAS ALSO DUE TO A LOOSE SHELL AND THE PATIENT HAD BONE DEFICIENCY. RECORDS ARE AVAILABLE FOR FURTHER REVIEW.

Description of Event or Problem · 1

LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, INFLAMMATION, DIFFICULTY AMBULATING, AND DISABILITY. IT IS ALSO ALLEGED THAT THE PATIENT FELT THE HIP MOVE OUT OF PLACE. AN X-RAY CONFIRMED THAT THE ACETABULAR COMPONENT MOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
53059 ARTICULEZE M HEAD 36MM +1.5 FEMORAL HEAD JDI DEPUY ORTHOPAEDICS INC US 2213913

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention