FDA Adverse Event
Malfunction
Summary report: N
AEZ LONG HANDLE MINI PIN & LIGATURE CUTTER
MDR report key: 2946742
·
Received February 6, 2013
Report
- Report Number
- 2016150-2013-00002
- Event Type
- Malfunction
- Date Received
- February 6, 2013
- Report Date
- November 1, 2012
- Manufacturer
- ORMCO CORPORATION
- Product Code
- EJB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PATIENT DETAILS WITH REGARD TO AGE, GENDER, AND WEIGHT WERE NOT PROVIDED. IT WAS CONFIRMED THAT NO SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT; THE BROKEN PIECE WAS COLLECTED IMMEDIATELY. THE PATIENT WAS NOT HARMED AND IS DOING FINE.A VISUAL TEST WAS PERFORMED ON THE RETURNED DEVICE CONFIRMING THAT THE CUTTING SURFACE HAD BROKEN AWAY FROM THE TIP AND A BRAZE FAILURE HAD OCCURRED; HOWEVER, THE CAUSE OF THE BREAKAGE WAS INCONCLUSIVE. IN ADDITION, NO SIMILAR COMPLAINTS WERE RECEIVED WITH REGARD TO THIS LOT NUMBER.
Description of Event or Problem · 1
A DOCTOR'S OFFICE ALLEGED THAT THE BLADE FROM AN AEZ MINI LIGATURE CUTTER HAD BROKEN OFF OF THE DEVICE DURING A PATIENT'S PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51247 | AEZ LONG HANDLE MINI PIN & LIGATURE CUTTER | LIGATURE WIRE CUTTER | EJB | ORMCO CORPORATION | 08H10H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |