FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2946371 · Received February 6, 2013

Report

Report Number
2955842-2013-00423
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
December 28, 2012
Report Date
January 11, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE AS THE INSTRUMENT PASSED RECOGNITION AND ENGAGEMENT TESTING WHEN INSTALLED ON AN IN-HOUSE IS3000 SYSTEM. THE INSTRUMENT MOVED INTUITIVELY WITH A FULL RANGE OF MOTION IN ALL DIRECTIONS AND THE GRIPS OPENED AND CLOSED PROPERLY. ENGINEERING EVALUATION ALSO FOUND THAT THE PITCH DOWN CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI DIAGNOSTIC LAPAROSCOPY PROCEDURE, THE JAW ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT STOPPED WORKING. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50907 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120820 702

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM, ACCESSORIES & ESU.