ENDOPATH XCEL BLUNT TIP TROCAR
Report
- Report Number
- 3005075853-2013-00525
- Event Type
- Malfunction
- Date Received
- February 6, 2013
- Date of Event
- January 16, 2013
- Report Date
- January 17, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). DOES THIS TROCAR HAVE THE OPTIVIEW TECHNOLOGY? ---NO. WERE ANY NOISES HEARD SUCH AS WHISTLING OR HISSING? ---YES. IF SO, DID THE NOISE PREVENT INSUFFLATION? ---YES PLEASE DESCRIBE THE NOISE. ---NO INFORMATION. WAS THERE A DROP IN PRESSURE? ---NO IF YES, DID THIS AFFECT THE VISIBILITY OF THE SURGEON? ---NO. WHAT WAS THE GAS CONSUMPTION RATE OR VOLUME (LITER/MINUTE)? ---EIGHT TO 10 L/MIN. WERE YOU ABLE TO IDENTIFY WHERE THE LEAK WAS COMING FROM? ---VALVE. WAS A DEVICE INSERTED IN THE TROCAR DURING THE LEAKING? ---YES. IF SO, WHAT DEVICE? ---CAMERA.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, AIR LEAKED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51009 | ENDOPATH XCEL BLUNT TIP TROCAR | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |