FDA Adverse Event Malfunction Summary report: N

ENDOPATH XCEL BLUNT TIP TROCAR

MDR report key: 2946205 · Received February 6, 2013

Report

Report Number
3005075853-2013-00525
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
January 16, 2013
Report Date
January 17, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). DOES THIS TROCAR HAVE THE OPTIVIEW TECHNOLOGY? ---NO. WERE ANY NOISES HEARD SUCH AS WHISTLING OR HISSING? ---YES. IF SO, DID THE NOISE PREVENT INSUFFLATION? ---YES PLEASE DESCRIBE THE NOISE. ---NO INFORMATION. WAS THERE A DROP IN PRESSURE? ---NO IF YES, DID THIS AFFECT THE VISIBILITY OF THE SURGEON? ---NO. WHAT WAS THE GAS CONSUMPTION RATE OR VOLUME (LITER/MINUTE)? ---EIGHT TO 10 L/MIN. WERE YOU ABLE TO IDENTIFY WHERE THE LEAK WAS COMING FROM? ---VALVE. WAS A DEVICE INSERTED IN THE TROCAR DURING THE LEAKING? ---YES. IF SO, WHAT DEVICE? ---CAMERA.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, AIR LEAKED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
51009 ENDOPATH XCEL BLUNT TIP TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1