FDA Adverse Event
Malfunction
Summary report: N
INFUSO.R.
MDR report key: 2946199
·
Received February 6, 2013
Report
- Report Number
- 1416980-2013-03070
- Event Type
- Malfunction
- Date Received
- February 6, 2013
- Date of Event
- August 31, 2012
- Report Date
- September 20, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K883577
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE REPORTED PROBLEM WAS CONFIRMED BY BAXTER QUALITY ENGINEERING AS MOTOR ISSUE. THE CAUSE OF THE PROBLEM WAS THE MOTOR BEING DISENGAGED FROM THE GEAR BOX AND THE RED BATTERY WIRE BEING BROKEN. NO REPAIRS WERE MADE AS THE PUMP WAS SENT FOR DESTRUCTION.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED AN INFUSOR PUMP WITH A CONDITION OF "PUMP IS NOT WORKING AT ALL. CUSTOMER WOULD LIKE AN "OVERHAUL" ON MACHINE". IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. HOWEVER, IT IS KNOWN TO HAVE OCCURRED IN THE "OUTPATIENT SURGERY" DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO PATIENT/USER INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE FACILITY REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51268 | INFUSO.R. | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|