FDA Adverse Event Malfunction Summary report: N

INFUSO.R.

MDR report key: 2946199 · Received February 6, 2013

Report

Report Number
1416980-2013-03070
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
August 31, 2012
Report Date
September 20, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED PROBLEM WAS CONFIRMED BY BAXTER QUALITY ENGINEERING AS MOTOR ISSUE. THE CAUSE OF THE PROBLEM WAS THE MOTOR BEING DISENGAGED FROM THE GEAR BOX AND THE RED BATTERY WIRE BEING BROKEN. NO REPAIRS WERE MADE AS THE PUMP WAS SENT FOR DESTRUCTION.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSOR PUMP WITH A CONDITION OF "PUMP IS NOT WORKING AT ALL. CUSTOMER WOULD LIKE AN "OVERHAUL" ON MACHINE". IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. HOWEVER, IT IS KNOWN TO HAVE OCCURRED IN THE "OUTPATIENT SURGERY" DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO PATIENT/USER INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE FACILITY REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
51268 INFUSO.R. PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1