FDA Adverse Event Injury Summary report: N

REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7

MDR report key: 2946176 · Received February 6, 2013

Report

Report Number
0002249697-2013-00576
Event Type
Injury
Date Received
February 6, 2013
Date of Event
January 10, 2013
Report Date
January 10, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K092561
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVENT REGARDING REVISION INVOLVING A REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. COMPLAINT HISTORY REVIEW: SIMILAR EVENTS HAVE OCCURRED FOR THE CATALOG NUMBER AND PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED REVISION IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL. NO FURTHER INVESTIGATION IS REQUIRED. NCR (B)(4) WAS INITIATED ON (B)(4) 2012 BECAUSE THE OCCURRENCE RATE FOR ADVERSE LOCAL TISSUE REACTION (ALTR) SPECIFIED IN THE RISK MANAGEMENT FILES FOR THE REJUVENATE MODULAR HIP SYSTEM WAS EXCEEDED. SEE CAPA (B)(4) FOR CORRECTIVE ACTIONS ASSOCIATED WITH THIS NCR. GIVEN THE POTENTIAL RISK OF FRETTING AND CORROSION WITH THESE DEVICES, VOLUNTARY PRODUCT RECALL RA 2012-067 WAS ISSUED.

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON REVISED PATIENT'S LEFT REJUVENATE HIP DUE TO CORROSION AT THE NECK STEM JUNCTION. PATIENT HAS LEGAL REPRESENTATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON REVISED PATIENT'S LEFT REJUVENATE HIP DUE TO CORROSION AT THE NECK STEM JUNCTION. PATIENT HAS LEGAL REPRESENTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50954 REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7 IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention