FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® COMPACT TEST DRUM

MDR report key: 2946147 · Received February 6, 2013

Report

Report Number
1823260-2013-00723
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
January 15, 2013
Report Date
March 6, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
K031755
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. THIS MEDWATCH IS FOR THE COMPACT PLUS SYSTEM 2. REFERENCE MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR THE COMPACT PLUS SYSTEM 1. DEVICE RECEIVED IN A CONDITION WHICH MADE ANALYSIS IMPOSSIBLE. UNABLE TO TEST BECAUSE USED STRIPS WERE RETURNED.

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. THIS MEDWATCH IS FOR THE COMPACT PLUS SYSTEM 2. REFERENCE MEDWATCH WITH (B)(6) FOR THE COMPACT PLUS SYSTEM 1.

Description of Event or Problem · 1

CUSTOMER ALLEGEDLY RECEIVED THE FOLLOWING RESULTS, WITH TWO DIFFERENT METERS, WITHIN 10 MINUTES: 13 MG/DL, 19 MG/DL (COMPACT PLUS SYSTEM 1) AND 130 MG/DL (COMPACT PLUS SYSTEM 2). NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND STRIPS, AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
49616 ACCU-CHEK ® COMPACT TEST DRUM BLOOD GLUCOSE MONITORING TEST STRIPS LFR ROCHE DIAGNOSTICS NA 20732941

Patients

Seq Age Sex Outcome Treatment
1 070 YR CALCIUM| ASPIRIN| SIMVASTATIN| GLUCOSAMINE| METFORMIN| LISINOPRIL