ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-01529
- Event Type
- Malfunction
- Date Received
- February 6, 2013
- Report Date
- January 28, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO DAMAGE WAS OBSERVED. DURING TESTING, ALL OF THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE BUTTON PRESSES WITH INCREASED FORCE TO ELICIT A RESPONSE. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER THE DOWN ARROW KEY CONTACT.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE UP ARROW, DOWN ARROW AND OK KEYPAD BUTTON HAD AN INTERMITTENT, DELAYED RESPONSE AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 50302 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |